Overview  
Throughput Rate 624 tests/hour: 437 tests/hour photometric, 187 tests/hour ISE
167 immunoassay tests/hour
Assay Capacity Onboard 47 including 3 ISE
   
Sample Handling  
Sample Tubes 5 mL, 7 mL, 10 mL tubes, 1.5 and 1.0 mL sample cups, pediatric tubes
Sample Area 60 samples in 6 x 10 tube segments, positive sample identification
Sample Integrity Check Qualitative check for hemolysis, lipemia, and icterus
STAT Sample Loading 60, not dedicated
Bar Codes 39, Code 128, Codabar (USS) Interleaved 2 of 5 with and without check digit
Auto-Repeat Automatic repeat testing from the original sample
Auto-Dilution Automatic dilution from the original sample
Auto-Reflex Testing Automatic ability based on results of first test
Primary Probe Liquid level sensing, clot check, short sample detection
   
Microvolume Technology  
Assay-Specific Autodilute Capability 1–1.6 up to 1–200, automatic urine dilutions
Original Sample Volume 2–60 uL per test
Average Reagent Volume 80–120 uL per test
Storage Capacity Onboard 12,600 tests average; 16,650 tests maximum
   
Reaction Area  
Reaction Tray 12,000 cuvettes formed onboard before reload
Path Length D1+T3 - [E1+T1+T2] = 0.5 cm +/- 0.0125 cm
Photometer The filter wheel holds the optical filters at 293 nm, 340 nm, 383 nm, 405 nm, 452 nm, 510 nm, 540 nm, 577 nm, 600 nm, 700 nm
Light Source Standard tungsten halogen lamp, operation at 6.5A (6.8v) lamp current
provides >1µA of zero absorbance measurement current at any wavelength
except 293 nm
Assay Methods Endpoint, rate, multipoint, immunoassay, turbidometric, LOCI
Reaction Times 3, 4, 5, 10, 15, and 21 minutes
Automatic Correction Serum blank, cell blank, reagent blank, measurement point change, autodilution
   
Reagent Handling  
Reagent Tray 1 tray, 44 positions
Reagent Capacity Onboard 44 Flex® Reagent Cartridges plus 3 electrolytes via the QuikLYTE® IMT
Dispensing System 2 probes with Liquid Level Sensing
Reagent Containers Flex® Reagent Cartridges, bar coded, 15 to 240 tests Flex
Reagent Inventory Management Tracks tests remaining, lot number, onboard stability, and expiration date
Onboard Stability Up to 30 days
   
Open System Capability  
Channels 10 channels, includes user-defined applications
Third-Party Applications Varies by country, can be configured on system
   
ISE  
ISE Indirect simultaneous measurement of Na+, K+, Cl-
Sample Volume 40 μL for all three tests
Priming Automatic priming cycle, no user calibration, automatic urine dilution 1:10
Electrode Shelf Life 1,000 samples or 3 days
Throughput Rate 187 tests/hour; 62 tubes/hour
   
Calibration/QC  
Calibration Interval Up to 90 days, tracked by software
Auto-Calibration/Auto-QC User-defined time interval or with new reagent container
View Calibration/QC Data Graphical display of calibration curves and QC; RealTime QC
   
Data Management  
Operating Computer Linux Operating System, 1 GB RAM, touch screen monitor 17”
System Documentation Operator manual and assaysheets online
Host Interface RS 232C bidirectional
Data Storage 100,000 patient tests (10 MB), 100,000 QC results (10 MB),
9,000 Calibrations (5 yrs, 18 MB)
Host Query ASCII; system requests work order or batch of work orders from host
   
General Specifications  
Water Requirements

 

  • Instrument is supplied with a water purifier that provides instrument feed water
  • Instrument feed water system must maintain stable d02 content between 5 and 8 ppm
  • Consumption 1.32 gal/hr (5.0L/hr) at maximum throughput
  • Temperature: <35°C
  • Resistivity: >10 megaohms cm
  • Bacterial content: < 10 colony forming units/mL
  • System feed water line must not exceed 12 feet
  • Water system may or may not be a part of the sale. Customers may purchase their own water system. A customer who purchases our Millipore system, must adhere to our water specifications.
Drain Requirements Maximum of 10.6 gallons (40 liters) per hour
Dimensions 56” W x 49” H x 41” D (without monitor)
Weight 770 lbs (349 kg)
Noise Specifications <75 dBA at 1 m while operating
Average Heat Output 3,753 BTU/hr
Operating Temperature Range 64º – 86ºF (18º – 30ºC)

 